Figure 3-32: Transmitter Electronics Enclosure removal
A. Transmitter electronics enclosure
B. Base enclosure bolts
C. Base enclosure
5. Use a 3 mm flat head screw driver and disconnect the Acquisition cable terminal
block and the transducer wire terminal blocks from the Acquisition Module inside of
the Base enclosure.
Meter repairs Maintenance and Troubleshooting manual
July 2020 3-9000-769
90 Gas Ultrasonic Flow Meters