Daniel
Series 1500 Liquid Turbine Meter
NPS 3 through 24

Table of Contents

Questions and Answers

  • L
    Lee WalkerJul 28, 2025
    How to fix inaccurate readout on Emerson Daniel Measuring Instruments?
    • J
      Jamie BarnesJul 28, 2025
      If you're experiencing an inaccurate readout, start by checking the rotor blades for any foreign material and clean them. If the rotor blades are bent, it's recommended to return the unit to the factory for replacement or repair. Also, troubleshoot any accessory equipment connected to the measuring instrument.
  • Y
    yriveraAug 1, 2025
    Why is the readout inaccurate on my Emerson Daniel Measuring Instruments?
    • W
      William HernandezAug 1, 2025
      An inaccurate readout on your Emerson Measuring Instruments can stem from a few potential issues. Check the rotor blades for any foreign material and clean them if necessary. If the rotor blades are bent, the rotor assembly should be returned to the factory for replacement or repair. Also, defective accessory equipment may cause an inaccurate readout, so troubleshoot that equipment.
  • M
    Matthew GriffithAug 7, 2025
    What to do if Emerson Daniel Measuring Instruments have no output pulses from amplifier module?
    • J
      Joshua WagnerAug 7, 2025
      If there are no output pulses from the amplifier module, several factors could be at play. First, ensure the input voltage to the amplifier is within the required range of 10-30 VDC. If the voltage is correct, check the pickoff for damage or shorts; its resistance across the leads should be between 600-900 Ohms, and replace it if necessary. If the pickoff is functional, the amplifier module itself may be damaged and require replacement. Also, troubleshoot the internal assembly. If the customer receiver unit is not operating, it is recommended to contact Emerson Flow customer service.
  • N
    Nicole RollinsAug 12, 2025
    Why is the turbine meter rotor not turning in my Emerson Daniel Measuring Instruments?
    • J
      Jean ReevesAug 12, 2025
      If the turbine meter rotor isn't turning, it could be due to a defective rotor bearing or damage from foreign material passing through the turbine meter. In either case, it's recommended to return the rotor assembly to the factory for replacement or repair.
  • T
    Todd NealAug 17, 2025
    What to do if there are no output pulses from the Emerson Daniel amplifier module?
    • J
      Justin GutierrezAug 17, 2025
      If you are experiencing no output pulses from the amplifier module, several factors could be the cause. First, ensure that the input voltage to the amplifier is within the required range of 10-30 VDC. If the voltage is correct, the pickoff might be damaged or shorted; in this case, replace the pickoff. A damaged amplifier module itself could also be the problem, requiring replacement. If the customer receiver unit is not operating, contact Daniel Customer Service. Lastly, if the meter rotor is not turning, troubleshoot the internal assembly.
  • J
    jdawsonAug 22, 2025
    What to do if the turbine meter rotor is not turning in my Emerson Daniel?
    • E
      egallagherAug 22, 2025
      If the turbine meter rotor isn't turning, it could be due to a defective rotor bearing or damage from foreign material. In either case, return the rotor assembly to the factory for replacement or repair.

Summary

Part I Plan

Introduction

Purpose, description, configurations, operation, parts lists, and certifications for the Series 1500 LTM.

Operating Conditions and Specifications

Details operating conditions, design factors, parameters, interface specs, and installation requirements.

Turbine Meter Handling

Covers receiving, storing, packing, and handling procedures for the turbine meter.

Prepare the Turbine Meter for Use

Discusses lifting, configuration, piping, calibration, grounding, and security seal installation.

Part II Install

Installation Prerequisites

Covers pre-start checks, torque information, and required tools for installation.

Installation Procedure

Details cathodic protection, mechanical assembly, and electronic component installation.

Testing

Outlines procedures for commissioning the turbine meter after installation.

Part III Operate

Operation Parameters

Covers operation overview, normal operation, back pressure calculation, and remote operation.

Part IV Maintain

Planned Maintenance

Details considerations, tools, disassembly, assembly, enclosures, preamplifier replacement, and tasks.

Corrective Maintenance

Covers troubleshooting and verifying operational condition after maintenance.

Spare Parts

Identifies recommended spare parts and provides ordering information.

Decommission the Turbine Meter

Details procedures for shutting down, disassembling, and shipping the meter.

Emerson Daniel Series Specifications

General IconGeneral
TypeFlow Meter
Repeatability±0.1% of reading
Communication ProtocolHART, Modbus, Foundation Fieldbus
Output4-20 mA, pulse, frequency
Process ConnectionFlanged, wafer, threaded
MaterialStainless steel, carbon steel, Hastelloy
Operating Temperature-40°C to +85°C (-40°F to +185°F)
Protection RatingIP67, NEMA 4X
Fluid CompatibilityLiquids, gases