Instruction Manual
D102748X012
DLC3010 Digital Level Controller
Introduction and Specifications
May 2018
10
Figure 1‐3. Theoretical Reversible Temperature Effect on Common Torque Tube Materials, Degrees Fahrenheit
TORQUE RATE AMPLIFICATION
(NORMALIZED MODULUS OF RIGIDITY)
TEMPERATURE (_F)
G
norm
1.10
1.08
1.06
1.04
1.02
1.01
1.00
1.05
1.03
1.09
1.07
-320 -280 -240 -200 -160 -120 -80 -40 0 40 80
TORQUE RATE REDUCTION
(NORMALIZED MODULUS OF RIGIDITY)
TEMPERATURE (_F)
N05500
N06600
N10276
S31600
50 100 150 200 250 300 350 400 450 500 550 600 650 700 750 800
1.00
0.96
0.92
0.88
0.84
0.82
0.80
0.90
0.86
0.98
0.94
G
norm
1
NOTE:
 1 DUE TO THE PERMANENT DRIFT THAT OCCURS NEAR AND ABOVE 500_F, N05500 IS NOT
RECOMMENDED FOR TEMPERATURES ABOVE 450_F.
 2 FOR PROCESS TEMPERATURES BELOW -20_F AND ABOVE 400_F SENSOR
MATERIALS MUST BE APPROPRIATE FOR THE PROCESS; SEE TABLE 1‐4.
CRYOGENIC
N05500
S31600

Table of Contents

Other manuals for Emerson FISHER FIELDVUE DLC3010

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson FISHER FIELDVUE DLC3010 and is the answer not in the manual?

Emerson FISHER FIELDVUE DLC3010 Specifications

General IconGeneral
BrandEmerson
ModelFISHER FIELDVUE DLC3010
CategoryController
LanguageEnglish

Related product manuals