Chapter3
SetuptheWeights&Measuresapplication
Topicscoveredinthischapter
♦Locale-speciccommissioning
InformationinthischapterisapplicableonlyifyourtransmitterwasorderedwiththeWeights&
Measuresapplication.
3.1Locale-speciccommissioning
3.1.1ReadFieldVericationZero(FVZ)
Display
ActivateSCROLLuntilFVZappears.
(1)
ProLinkII
ProLink→DiagnosticInformation
Communicator
Notavailable
TheFieldVericationZero(FVZ)diagnosticvariableisreadduringmetercommissioningtocomply
withMIDrequirementsforWeights&Measuresapplications.
3.1.2Readthermwarechecksum
Display
OFF-LINEMAINT→SWREV
ProLinkII
ProLink→Conguration→Device→FirmwareChecksum
ProLink→Conguration→Device→CPFirmwareChecksum
ProLink→CoreProcessorDiagnostics
Communicator
5,4,TransmitterFirmware
DetailedSetup→DeviceInformation→TransmitterFirmware
5,4,CoreProcessorFirmware
DetailedSetup→DeviceInformation→CoreProcessorFirmware
Thechecksumvaluesforthetransmitterrmwareandthecoreprocessorrmwarearereadduring
metercommissioningtomeetWeights&MeasuresrequirementsforgasapplicationsinGermany.
TheymayalsobeusefulforMID/Welmec7.2testreports.
(1)ToreadFVZfromthedisplay,itmustbeconguredasadisplayvariable.
CongurationandUseManualSupplement