Reference Manual
00809-0100-4690, Rev FC
June 2011
Rosemount 2088 and 2090
5-6
Device write protected. Do you
still want to
shut off?
Device is in write-protect mode. Press YES to turn the
Field communicator off and lose the unsent data.
Display value of variable on
hotkey menu?
Asks whether the value of the variable should be
displayed adjacent to its label on the hotkey menu if the
item being added to the hotkey menu is a variable.
Download data from
configuration memory to device
Press the SEND softkey to transfer information from the
communicator memory to the device.
Exceed field width Indicates that the field width for the current arithmetic
variable exceeds the device-specified description edit
format.
Exceed precision Indicates that the precision for the current arithmetic
variable exceeds the device-specified description edit
format.
Ignore next 50 occurrences of
status?
Select YES to ignore the next 50 occurrences of device
status, or select no to display every occurrence.
Illegal character An invalid character for the variable type was entered.
Illegal date The day portion of the date is invalid.
Illegal month The month portion of the date is invalid.
Illegal year The year portion of the date is invalid.
Incompatible CPU board and
module ON
Upgrade the electronics board or the sensor module to
the current revision.
Incomplete exponent The exponent of a scientific notation floating point
variable is incomplete.
Incomplete field The value entered is not complete for the variable type.
Looking for a device Polling for multidropped devices at addresses 1–15.
Local buttons operator error ON Illegal pressure applied during zero or span operation.
Repeat the process after verifying the correct pressures.
Mark as read only variable on
hotkey menu?
Asks whether the user should be allowed to edit the
variable from the hotkey menu if the item being added to
the hotkey menu is a variable.
Module EEPROM write failure
ON
Message sent to the module from the HART signal failed.
Replace the transmitter.
No device configuration in
configuration memory
There is no configuration saved in memory available to
re-configure off-line or transfer to a device.
No Device Found Poll of address zero fails to find a device, or poll of all
addresses fails to find a device if auto-poll is enabled.
No hotkey menu available for
this device.
There is no menu named “hotkey” defined in the device
description for this device.
No pressure updates ON No pressure updates being received from the sensor
module. Verify that the sensor module ribbon cable is
attached correctly. Or replace the transmitter.
No offline devices available. There are no device descriptions available to be used to
configure a device offline.
No simulation devices available. There are no device descriptions available to simulate a
device.
No temperature updates ON No temperature updates being received from the sensor
module. Verify that the sensor module ribbon cable is
attached correctly. Or replace the transmitter.
No UPLOAD_VARIABLES in ddl
for this device
There is no menu named “upload_variables” defined in
the device description for this device. This menu is
required for offline configuration.
No Valid Items The selected menu or edit display contains no valid items.
OFF KEY DISABLED Appears when the user attempts to turn the HC off before
sending modified data or before completing a method.
Online device disconnected
with unsent data. RETRY or OK
to lose data.
There is unsent data for a previously connected device.
Press RETRY to send data, or press OK to disconnect
and lose unsent data.
Message Description