5.6.1 Use the integral buttons to change the level at 4mA
Procedure
1. From the PV display, press the green button ↓ repeatedly until 4 is
shown.
2. Press the blue button → to indicate the present value of the 4 mA
level.
• If this value is correct, press the red button ↵ and then the green
button ↓ to get to the next menu.
• Otherwise, continue with step 3.
3. Press the blue button → to start editing. The first digit flashes.
4. Press the green button ↓ repeatedly to edit the flashing digit.
5. Press the blue button → to move to the next digit. That digit flashes.
6. Repeat steps 5 and 6 until the last digit is flashing and edited as
required.
7. Press the blue button → to confirm the new 4 mA level. No flashing
digits.
• Press the red button ↵ to save the new 4 mA level. The next
menu appears.
• Press the blue button → to exit without saving. The 4 menu
appears.
5.6.2 Use the integral buttons to change the level at 20mA
Procedure
1. From the PV display, press the green button ↓ repeatedly until 20 is
shown.
2. Press the blue button → to indicate the present value of the 20 mA
level.
• If this value is correct, press the red button ↵ and then the green
button ↓ to get to the next menu.
• Otherwise, continue with step 3.
3. Press the blue button → to start editing. The first digit flashes.
4. Press the green button ↓ repeatedly to edit the flashing digit.
5. Press the blue button → to move to the next digit. That digit flashes.
6. Repeat steps 5 and 6 until the last digit is flashing and edited as
required.
August 2019 Quick Start Guide
Quick Start Guide 33