August 2015
8
Quick Start Guide
Figure 5. Sensor Orientation
3.4 Electrode orientation
The electrodes in the sensor are properly oriented when the two measurement
electrodes are in the 3 and 9 o’clock positions or within 45 degrees from the
horizontal, as shown on the left of Figure 6. Avoid any mounting orientation that
positions the top of the sensor at 90 degrees from the vertical position as shown
in Figure 6.
Figure 6. Mounting Position
For hazardous location installations, refer to Appendix D of Reference Manual
00809-0400-4444 for sensor orientation pertaining to specific T-code
compliance.