8.2.3  Reinstalling or replacing the SAM42 detector
To remove the detector unit from the pipe, unscrew the four
mounting nuts of the bolts using a 13 mm spanner, being careful
to avoid nuts and springs falling off, then pull up the detector from
the mounting socket, sliding through the bolts.
Replacing a SAM42 detector will require a new calibration. Before
installing the new detector, ensure it can be done safely. The
detector’s weight is ~3kg and it can cause damage if accidentally
dropped. Use scaffolding if required.
To physically install the new detector, follow the steps detailed in
Physical installation of the device.