3.5 Electrode orientation
The electrodes in the sensor are properly oriented when the two
measurement electrodes are in the 3 and 9 o’clock positions or within 45
degrees from the horizontal, as shown on the left side of Figure 3-4. Avoid
any mounting orientation that positions the top of the sensor at 90 degrees
from the vertical position as shown on the right of Figure 3-4.
Figure 3-4: Electrode orientation
A
B
A. Correct orientation
B. Incorrect orientation
The sensor may require a specific orientation to comply with Hazardous Area
T-code rating. Refer to the appropriate reference manual for any potential
restrictions.
November 2019 Quick Install Guide
Quick Install Guide 11