TROUBLESHOOTING
Symptom Solution
Power setting LEDs do not illuminate
when the unit is switched on
Check all the connections
Check the power supply of the dental office
Send the unit to an approved EMS repair center
No or insufficient irrigation flow when
the foot pedal is pressed
Turn the liquid flow rate knob
Make sure that the "DRY WORK" mode is not activated
Make sure that the bottle is properly installed
Check all the connections
Proceed to a circuit washing without the handpiece and check that liquid exits the handpiece cord; if it is the
case, the handpiece is plugged; send the handpiece to an approved EMS repair center
Unscrew the pump cover and check that the peristaltic pump functions; if the rollers do not turn, replace the
pump (the rotation of the rollers can be seen by transparency)
Send the unit to an approved EMS repair center
No ultrasound oscillations or
insufficient power
Control that the instrument is tight
Check the wear of the instrument and replace it using the CombiTorque if necessary
Press the foot pedal to the second step
Check all the connections
Dry the connectors of the handpiece cord and of the handpiece
Make sure that the "ENDO" mode is not activated
Increase the power of the ultrasound to the maximum
Replace the handpiece cord and/or the handpiece
Send the unit to an approved EMS repair center
Water leaks Replace the O-ring on the bottle screw-cap
Replace the O-ring on the handpiece and on the handpiece cord
Replace the bottle
Replace the peristaltic pump
Send the handpiece and the handpiece cord to an approved EMS repair center
Insufficient lighting of the Piezon
handpiece LED
Replace the light guide
Send the unit to an approved EMS repair center