6. Turn Unit O:
by the plug, not the cord. If treatment will be resumed shortly the electrodes may be left on the skin.
When the electrodes are removed, clean the skin thoroughly with mild soap and water. If there is
skin irritation, consult your medical professional.
Caution: When the timer runs out, the unit will turn o automatically and you are not required to
press the “ON/OFF” button. Unit will also turn o automatically after 5 minutes if no activity is
made when the unit is initially turned on.
NOTE:
is device is safeguarded with a Locking mechanism to avoid possible mishandling when treatment is in session.
e user can manually “Lock” the settings for the duration of the treatment by pressing and holding the “key” button for 3
seconds (a beep (if enabled) will conrm device is locked). e device also automatically “locks”if no treatment button is
pressed for 1 minute. Once locked, other than the power button, no buttons are active. To disable the “Lock” during
treatment session for adjustment or stoppage, press and hold the “key” button for 3 seconds (a beep will conrm
device is unlocked). e user can then adjust amplitude, settings, etc.
NOTE: is device is capable of “locking” out either TENS or EMS features. To isolate device into a specic treatment
(EMS or TENS) mode, ensure device is not providing any stimulation and is on the desired treatment mode. Press and
hold Channel 1 and Channel 2 negative (-) amplitude button for 5
-
7 sec until audible beep is heard. is will allow only
TENS or only EMS modes to function. To reverse, follow above steps.