EasyManua.ls Logo

EMSI FLEX-MT PLUS - Declarations-EMC; Electromagnetic Emissions Guidance

EMSI FLEX-MT PLUS
42 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
28
RF emissions
CISPR 11
Group 1
RF emissions
CISPR 11
Class B
Harmonic
emissions
IEC 61000-3-2
Class A
Voltage
uctuations
/icker emissions
IEC 61000-3-3
Compliance
15.0 | Declarations-EMC
Guidance and manufacturer’s declaration-electromagnetic emissions
e Flex-MT® Plus is intended for use in the electromagnetic environment specied below.
e customer or the user of the Flex-MT® Plus should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment-guidance
e Flex-MT® Plus is suitable for use in all establishments,
including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
e Flex-MT® Plus uses RF energy only for its internal
function. erefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.