Electrostatic
discharge(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Electrical fast
transient/burst
IEC 61000-4-4
±2kV for power
supply lines
±1kV for input
output lines
±2kV for power
supply lines
Not applicable
+ 1kV differential
mode not
applicable
Surge IE
61000-4-5
±1kV line(s)
to line(s)
±2kV line(s)
to earth
Guidance and manufacturer’s declaration-electromagnetic immunity
e Flex-MT® Plus is intended for use in the electromagnetic environment specied below.
e customer or the user of the Flex-MT® Plus should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment-guidance
29
Floors should be wood, concrete or ceramic
tile. If oors are covered with synthetic
material, the relative humidity should be at
least 30%
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment.