Voltage Dips,
short interruptions
and voltage
variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic eld
IEC 61000-4-8
e Flex-MT® Plus power frequency
magnetic elds should be at levels
characteristic of a typical location in a
typical commercial or hospital environment.
3A/m
NOTE-UT is the a.c. mains voltage prior to application of the test level.
3A/m
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment-guidance
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the Flex-MT® Plus
requires continued operation during power
mains interruptions, it is recommended that
the Flex-MT® Plus be powered from an
uninterruptible power supply or a battery.
<5% UT(>95%
dip in UT) for
0.5 cycle
40% UT(60%
dip in UT) for 5
cycles
70% UT(30%
dip in UT) for
25 cycles
<5% UT(>95%
dip in UT) for 5
s
<5% UT(>95%
dip in UT) for
0.5 cycle
40% UT(60%
dip in UT) for 5
cycles
70% UT(30%
dip in UT) for
25 cycles
<5% UT(>95%
dip in UT) for 5
s
Guidance and manufacturer’s declaration-electromagnetic immunity continued