Operating TOPwave
30
Selects an application and push [OK]. The Program menu window appears.
Select the vessel for the digestion.
Master login only
When required change the parameter settings at your own applications.
A program always consists of 5 steps. The following parameters are displayed or can be
changed:
Parameter Description
TEMPERATURE Display and input of the target temperature in °C*.
Permitted values: 50-300°C, configurable in the setup
PRESSURE Display and input of the maximum permissible pressure in bar*. Permit-
ted values: 0-130 bar, configurable in the setup
RAMP Display and input of the length of time in minutes required to heat to the
target temperature. Permitted values: 0-59 min.
TIME Display and input of the length of time in minutes for which the target
temperature is to be kept constant. Permitted values: 0-59 min.
POWER Display and input of the maximum permissible power output in % at
which heating is to be maintained in this step of the program. Permitted
values: 0-100 %, configurable in the setup
INT Integral part of the PI controller
P-BAND Proportional gain of the PI controller
TOTAL TIME Display of the total runtime in minutes.
The maximum permissible runtime of a program is defined in the setup.
The permissible runtime is only queried during digestion; if this is excee-
ded the digestion is aborted.
VESSEL Number of vessels. This is updated when the type of vessel is selected.