TOPwave Operating
31
MAGT. Max. permissible magnetron temperature. If this temperature is exceeded
the digestion is aborted.
Permitted values: 0 -120°C
Typical values: 80 90°C
Values <20°C are used only for the venting programs.
Push [
ADD] to accept the application. Select [SAVE+EXIT], if you modified the applica-
tion.
The Main menu appears. The application is activated and can be used for the next di-
gestion.
Note:
Only user with
MASTER login can modify application. Standard applications cannot be
modified.
Select an application in the library and push [
MODIFY].
Modify the parameters (see above).
With [
SAVE+ EXIT] confirm the changes and return into the Main menu. The changed
application is used for the next digestion.
Of the 200 applications in the selective list 30 are pre-installed and 25 are currently al-
located.
Program number Description
1 Calcite, Dolomite
2 Cleaning, Vessel conditioning
3 Coal, DIN 22022-1
4 Complexation H3BO3
5 Evaporation
6 Feed
7 Feldspar
8 Filter, glass fiber
9 Fly Ash, DIN EN 14385
10 Food (no/low fat)
11 Food (High fat)
12 Glass, Quartz
13 Limestone
14 Oil, Lubricant
15 Plastic, PE, ROHs
16 Plastic, Waste DIN 22022-1
17 Soil ISO 11466, EN 13346
18 Soil EPA 3051
19 Soil, Sediment EPA 3052
20 Tissue, Blood
21 Vegetables, Leaves
22 Vegetable Oil
Modifying application in
the library
List of pre-installed stan-
dard programs
1 25