Operating TOPwave
36
The input window appears in which the program can again be modified (but only
with master authorization) and a file name can be selected. The index is always au-
tomatically increased by 1 where the file name is the same.
NOTICE
If the program is changed in this op
erating step the change has an effect only on the
current digestion and not on the saved application.
Check the selected type of vessels and edit the number of used vessels ([
NUMBER]).
For changing file name press the file name button (in above figure [C
ALIBRATIONS_]).
The
SELECT FILE window appears.
In this window you can select a file name, create a new file or reset the index num-
ber.
For selecting a file mark the correspondent line in the list and press [
NEXT].
For creating a new file enter the file name in the
EDIT field, press [OK] and then
[
NEXT].
Push [
RESET] for resetting the index of marked file. With [RESET ALL] you reset the in-
dexes of all files. If you reset the index, all entries in the file will be deleted.
Start the digestion program with [
START DIGESTION].
The heating program starts after an initial period of approx. 20 s. During digestion
the highest sample temperature and pressure curve are displayed. Also all tempera-
tures inside the vessels are shown in the form of a bar graph.
Push [
STOP] to abort the program.
The data will be saved up to the time of program abort.