EasyManua.ls Logo

Endress+Hauser Cerabar S PMC71 - Page 83

Endress+Hauser Cerabar S PMC71
238 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cerabar S/Deltabar S/Deltapilot S FOUNDATION Fieldbus Parameter description (FF configuration program)
Endress+Hauser 83
Unit Flow/
STD_FLOW_UNIT
Entry
Index: 16
Data type: UNSIGNED16
Access: read, write, OOS
Select flow unit.
Prerequisite:
• Deltabar S differential pressure transmitter
!
Note!
Make sure that the unit suits the flow mode selected. → ä 82, parameter description for
Flow-Meas. Type/FLOW_MEAS_TYPE.
When a new flow unit is selected, all flow-specific parameters are converted and displayed
with the new unit within a flow mode (Flow-Meas. Type/FLOW_MEAS_TYPE). When the
flow mode is changed, conversion is not possible.
Possible units for Flow-Meas. Type/FLOW_MEAS_TYPE = Volume operat. cond.:
• m3/s, m3/min, m3/h, m3/day
• l/s, l/min, l/h
• hl/s, hl/min, hl/day
• ft3/s, ft3/min, ft3/h, ft3/day
• ACFS, ACFM, ACFH, ACFD
•ozf/s, ozf/min
• US Gal/S, US Gal/min, US Gal/h, US Gal/day
• Imp. Gal/s, Imp. Gal/min, Imp. Gal/h
• bbl/s, bbl/min, bbl/h, bbl/day
• User unit, see also parameter description for Customer Unit F/CUSTOMER_UNIT_
FLOW (→ ä 84) and Cust. Unit Fact. F/CUSTOMER_UNIT_FACTOR_FLOW (→ ä 84)
Factory setting:
m³/s
Possible units for Flow-Meas. Type/FLOW_MEAS_TYPE = Volume norm. cond.:
• NM3/s, Nm3/min, Nm3/h, Nm3/day
•User unit,
→ see also parameter description for Customer Unit F/CUSTOMER_UNIT_
FLOW (→ ä 84) and Cust. Unit Fact. F/CUSTOMER_UNIT_FACTOR_FLOW (→ ä 84)
Factory setting:
Nm³/s
Possible units for Flow-Meas. Type/FLOW_MEAS_TYPE = Volume std. cond.:
• Sm3/s, Sm3/min, Sm3/h, Sm3/day
• SCFS, SCFM, SCFH, SCFD
•User unit,
→ see also parameter description for Customer Unit F/CUSTOMER_UNIT_
FLOW (→ ä 84) and FCust. Unit Fact. F/CUSTOMER_UNIT_FACTOR_FLOW
(→ ä 84)
Factory setting:
Sm3/s
Possible units for Flow-Meas. Type/FLOW_MEAS_TYPE = Mass p. cond.:
• g/s, kg/s, kg/min, kg/h
• t/s, t/min, t/h, t/day
•oz/s, oz/min
• lb/s, lb/min, lb/h
• ton/s, ton/min, ton/h, ton/day
•User unit,
→ see also parameter description for Customer Unit F/CUSTOMER_UNIT_
FLOW (→ ä 84) and Cust. Unit Fact. F/CUSTOMER_UNIT_FACTOR_FLOW (→ ä 84)
Factory setting:
kg/s
DP Flow Transducer Block
Parameter Description

Table of Contents