Operation Cleanfit CPA871
44 Endress+Hauser
7 Operation
7.1 Adapting the assembly to the process conditions
L
CAUTION
Due to the operating principle, there is a connection between the process and the service
chamber. The service chamber can be pressurized as a result.
Process medium may escape during insertion/retraction.
‣
Check that the process seal is intact.
‣
Pipe-fit the rinse chamber outlet accordingly.
‣
Seal the rinse connections with dummy plugs.
A0023307
29 Position markings (service position)
Assembly with pneumatic drive
The assembly with pneumatic drive does not have any operating elements.