Cleanfit CPA871 Maintenance
Endress+Hauser 63
Seal replacement in the rinse connection adapter
A0030292
43 Replacing seals, Part 3
4 Lock nut AF19
5 Rinse connection adapter AF17
1. Release the lock nuts (pos. 4) using a 19 mm open-ended wrench or socket wrench (in
protective cover).
2. Unscrew the two rinse connection adapters (pos. 5).
A0030315
44 Replacing seals, Part 3
10 O-rings, rinse connection adapter
3. Remove the O-rings indicated (pos. 10).
4. Apply a thin layer of grease (e.g., Klüber Paraliq GTE 703) to the new O-rings.
5. Insert the O-rings into the corresponding grooves.