EasyManua.ls Logo

Endress+Hauser Cleanfit CPA871 - Apply a Thin Layer of Grease (E.g., Klüber Paraliq GTE 703) to the New O-Rings; O-Rings, Rinse Connection Adapter; Remove the O-Rings Indicated (Pos. 10)

Endress+Hauser Cleanfit CPA871
92 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cleanfit CPA871 Maintenance
Endress+Hauser 63
Seal replacement in the rinse connection adapter
4
5
4
5
A0030292
 43 Replacing seals, Part 3
4 Lock nut AF19
5 Rinse connection adapter AF17
1. Release the lock nuts (pos. 4) using a 19 mm open-ended wrench or socket wrench (in
protective cover).
2. Unscrew the two rinse connection adapters (pos. 5).
10
10
A0030315
 44 Replacing seals, Part 3
10 O-rings, rinse connection adapter
3. Remove the O-rings indicated (pos. 10).
4. Apply a thin layer of grease (e.g., Klüber Paraliq GTE 703) to the new O-rings.
5. Insert the O-rings into the corresponding grooves.