EasyManua.ls Logo

Endress+Hauser Cleanfit CPA871 - Lock Nut

Endress+Hauser Cleanfit CPA871
92 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance Cleanfit CPA871
56 Endress+Hauser
Seal replacement in the rinse connection
4
5
4
5
A0030292
 35 Replacing seals, Part 3
4 Lock nut
5 Rinse connection adapter
1. Release the lock nuts (pos. 4) using an open-ended wrench or socket wrench (AF 19
mm, in protective cover).
2. Unscrew the two rinse connection adapters (pos. 5) using an open-ended wrench or
socket wrench (AF 17 mm, in protective cover).
10
10
A0030315
 36 Replacing seals, Part 3
10 O-rings, rinse connection adapter
3. Remove the O-rings indicated (pos. 10).
4. Apply a thin layer of grease (e.g., Klüber Paraliq GTE 703) to the new O-rings.
5. Insert the O-rings into the corresponding grooves.