Cleanfit CPA871 Maintenance
Endress+Hauser 55
2
3
A0030291
 34 Replacing seals, Part 2
2 Process connection
3 O-ring in process connection
2. Remove the process connection (pos. 2).
3. Remove the O-ring (pos. 3) from the process connection (gasket).
4. Apply a thin layer of grease to the new O-ring (e.g., Klüber Paraliq GTE 703).
5. Insert the O-ring into the process connection.