Cleanfit CPA871 Operation
Endress+Hauser 47
Use a pneumatic pilot valve (4/2-way or 5/2-way) to insert/retract the assembly.
‣
Connect the two inputs.
If only one input is connected (e.g. for test purposes), the piston is blocked as the
sensor guide moves before the limit position lock is disabled.
Inserting/retracting the assembly if the compressed air supply fails
A0030306
32 Failure of compressed air supply
1 Limit position lock for service position
2 Limit position lock for measuring position
L
CAUTION
Risk of injury due to high medium pressure
‣
Depressurize the system.
If the compressed air supply fails, you can still move the assembly manually. Proceed as
follows here:
1. Use an open-ended wrench (AF 17 mm) to unscrew both limit position locks (item 1
and 2).
2. Move the assembly to the desired position.
3. Screw the limit position lock back in.