Maintenance Cleanfit CPA871
68 Endress+Hauser
8.2.5 Seal accessories
Seals, extension and ball valve
A0043794
1 Sealing plug
2 Flat seal
3 O-ring
4 Extension
5 Ball valve
1. Apply a thin layer of grease to the O-ring (3) of the rinse chamber and flat seal (2) of
the extension (4), (e.g. Klüber Paraliq GTE 703).
2. Place the O-ring (3) and flat seal (2) in the corresponding grooves of the rinse
connections.
3. Mount the sealing plug (1) and extension (4).
4. Seal the ball valve (5) with a Teflon tape.
5. Screw the ball valve (5) into the extension (4).