Commissioning CM14
20 Endress+Hauser
Parameter Configuration options Description
Inactive time 1 to 240 min
60 min
Duration of the process check.
Analog outputs Settings for analog outputs.
Current range 4-20 mA
0-20 mA
Current span for analog output.
Out 1 0/4 mA Numerical value 0.000 -
99999
0.0 pH
Physical value corresponding to the lower
range limit of the analog output.
Out 1 20 mA Numerical value 0.000 -
99999
12 pH
Physical value corresponding to the upper
range limit of the analog output.
Out 2 0/4 mA Numerical value–50 to 250 °C
0 °C
Temperature corresponding to the measuring
range lower limit of the temperature input.
Out 2 20 mA Numerical value–50 to 250 °C
100 °C
Temperature corresponding to the measuring
range upper limit of the temperature input.
Damping main value 0 to 60 s
0 s
Configure the damping for the low-pass
filtering of the input signals.
Relay 1/2 Settings for the relay outputs.
Function Off, Min limit, Max limit, In
band, Out band, Error
Configure the function of the relay.
No additional settings are possible in the event
of function = Error.
Assignment Main, Temp Assign the relay to the main or temperature
input.
Set point Numerical value
0.0
Configures the limit value.
Set point 2 Numerical value
0.0
Only for function In band or Out band.
Hyst. Numerical value
0.0
Configures the hysteresis.
Delay time 0 to 60 s
0 s
Configure the delay until the relay switches.
Factory default Reset the device settings to the factory settings.
Please confirm no, yes Confirm reset.
6.4.1 Configuration of the relays
The device has two relays with limit values that are either switched off or can be allocated to
the input signal. The limit value is entered as a numerical value including the position of the
decimal point. The way the changeover contact is wired ((→  35)) determines whether
the relays act as an NO or NC relay. Limit values are always assigned to a relay. Each relay can
be assigned a channel or calculated value. In "Error" mode, the relay acts as an alarm relay and
switches for each error or alarm.