Commissioning CPY7B
10 Endress+Hauser
5.2 Venting
Having topped up the electrolyte, you have to vent the electrolyte hose.
L
CAUTION
Minor skin or eye injuries are possible as a result of contact with the electrolyte
Wear safety gloves and protective goggles.
1. Release the electrolyte hose connection on the sensor.
2. Press in the tappet of the hose union until there is a flow of electrolyte:
A0005761
3. Reconnect the electrolyte hose to the sensor.
5.3 Assembly
Follow the sequence for filling the vessel with electrolyte in the reverse order.
1. Fit the top part and tighten the coupling nut (finger-tight, maximum 5 Nm).
2. Close the manually operated valve.
3. Connect the hose union.
4. Open the manually operated valve.