Deltabar FMD71, FMD72 Description of device parameters
47
Navigation Expert → Measurement → Sensor Trim HP → Hi Trim Measured (130)
Write permission No write permissions. Parameter is read only.
Description Displays the serial number of the sensor module HP (11 alphanumeric characters).
Lo trim sensor (131)
Navigation Expert → Measurement → Sensor Trim HP → Lo Trim Sensor (131)
Write permission No write permissions. Parameter is read only.
Description Sensor module recalibration by entering a target pressure while simultaneously and
automatically accepting a reference pressure present for the lower calibration point.
Hi trim sensor (132)
Navigation Expert → Measurement → Sensor Trim HP → Hi Trim Sensor (132)
Write permission No write permissions. Parameter is read only.
Description Sensor module recalibration by entering a target pressure while simultaneously and
automatically accepting a reference pressure present for the upper calibration point.
Start current (134)
Navigation Expert → Output → Current output → Startcurrent (134)
Write permission Service engineers/Expert
Description Entry of the start current. This setting also applies in HART multidrop mode.
Options • 12 mA
• Max Alarm (22 mA, cannot be adjusted)
Factory setting 12 mA
Curr. trim 4mA (135)
Navigation Expert → Output → Current output → Curr. trim 4mA (135)
Write permission Service engineers/Expert