Deltabar FMD71, FMD72 Description of device parameters
63
Write permission No write permissions. Parameter is read only.
Description Displays the measured pressure before the sensor trim.
Meas. press. HP (281)
Navigation Expert → Measurement → Pressure → Meas. Press. HP (281)
Write permission No write permissions. Parameter is read only.
Description Displays the measured HP pressure after sensor trim and simulation.
Meas. press. LP (282)
Navigation Expert → Measurement → Pressure → Meas. Press. LP (282)
Write permission No write permissions. Parameter is read only.
Description Displays the measured LP pressure after sensor trim and simulation.
Sensor Temp. LP (283)
Navigation Expert → Measurement → Basic setup → Sensor Temp. LP (283)
Write permission No write permissions. Parameter is read only.
Description Displays the temperature currently measured in the sensor module. This can deviate from
the process temperature.
Sim. Press. HP (284)
Navigation Expert → Diagnosis → Simulation → Sim. Press. HP (284)
Write permission Operators/Service engineers/Expert
Description Enter the simulation value. See also "Simulation Mode".
Prerequisite "Simulation Mode" = Pressure HP
Value at switch-on Current pressure measured value