Deltabar S PROFIBUS PA Installation
Endress+Hauser 11
In the case of level measurement in closed tanks with superimposed vapour, a condensate trap
ensures pressure which remains constant on the minus side.
Measuring arrangement level measurement in open tanks
PMD70, PMD75: Mount device below the lower measuring connection. The negative side is
open to atmosphere pressure.
FMD77: Mount device direct on the tank. The negative side is open to atmosphere pressure.
4.3.3 Flow measurement
The PMD75 are best suited for flow measurement.
Measuring arrangement for gases: Mount device above the measuring point.
Measuring arrangement for liquids and vapours: Mount device below tapping point.
For flow measurement in vapours, mount the condensate traps at the same level as the same
the tapping point and at the same distance from Deltabar S.