Operation Deltapilot M 4...20 mA HART
32 Endress+Hauser
Operating example: Accepting the pressure present
Example: setting position adjustment
6.3.4 Operation via SFX100
See operating instructions.
6.3.5 Operation via FieldCare
See operating instructions.
6.3.6 Locking/unlocking operation
See operating instructions.
6.3.7 Resetting to factory settings (reset)
See operating instructions.
The new value for the upper range value is 50.0 mbar (0.75 psi).
You exit the edit mode for the parameter with .
You can get back to the editing mode with or .
65 .000 mbar
Pos. zero adjust 007 Operation
1 Abort The pressure for position adjustment is present at the device.
Confirm
2 Confirm Use or to switch to the "Confirm" option. The active option is
highlighted in black.
Abort
3Compensation
accepted!
Accept the pressure present as position adjustment with the
key. The device confirms the adjustment and goes back to the
"Pos. zero adjust" parameter.
4 Abort Exit the edit mode for the parameter with .
Confirm
Set URV 014 Operation