EA01210P
Endress+Hauser 3
Only open housing for a brief period. Avoid the penetration
of foreign bodies, moisture or contaminants.
Replace defective seal/gaskets with genuine parts from
Endress+Hauser only.
If threads are damaged or defective, the measuring device
must be repaired.
Threads (e.g. of the cover for the electronics and connection
compartments) must be lubricated. Use an acid-free, non-
hardening grease if an abrasion resistant dry lubricant is
non-existent.
If spacing is reduced or the dielectric strength of the
measuring device cannot be guaranteed during repair work,
perform a test on completion of the work (e.g. high-voltage
test in accordance with the manufacturer's instructions).
Service connector:
Do not connect in potentially explosive atmospheres.
Only connect to Endress+Hauser service devices.
Observe the instructions for transporting and returning the
device outlined in the Operating Instructions.
If you have any questions, contact your
Endress+Hauser service organization.
Tools list
For HistoROM
(optional)
Replacing electronics
The HistoROM module (illustration 6) is optional.
1 →
2 → 3 → 4 →
2.5 Nm
3 mm
1.
2.
5 → 6
Re-assembly has to be done in reverse order.