Levelflex FMP50/51/52/53/54/55/56/57 XA00634F-A
6 Endress+Hauser
Safety instructions:
General
Staff must meet the following conditions for mounting, electrical installation, commissioning and
maintenance of the device:
- Be suitably qualified for their role and the tasks they perform
- Be trained in explosion protection
- Be familiar with national regulations
For installation, use and maintenance of the device, users must also observe the requirements stated in the
Operating Instructions and the standards:
- GB50257-1996: "Code for construction and acceptance of electric device for explosion atmospheres and
fire hazard electrical equipment installation engineering".
- GB3836.13-1997: "Electrical apparatus for explosive gas atmospheres, Part 13: Repair and overhaul for
apparatus used in explosive gas atmospheres".
- GB3836.15-2000: "Electrical apparatus for explosive gas atmospheres, Part 15: Electrical installations in
hazardous area (other than mines)".
- GB3836.16-2006: "Electrical apparatus for explosive gas atmospheres, Part 16: Inspection and
maintenance of electrical installation (other than mines)".
Install the device according to the manufacturer's instructions and national regulations.
Do not operate the device outside the specified electrical, thermal and mechanical parameters.
Only use the device in media to which the wetted materials have sufficient durability.
Refer to the temperature tables for the relationship between the permitted ambient temperature for the
sensor and/or transmitter, depending on the range of application, and the temperature class.
Modifications to the device can affect the explosion protection and must be carried out by staff authorized
to perform such work by Endress+Hauser.
When replacing the probe electronics or opening the connection between the remote cable and the probe,
a jumper plug must be used or a short-circuit must be established between the probe contact and the
potential equalization conductor to avoid electrostatically charging the probe.
Safety instructions:
Special conditions
Permitted ambient temperature range at the electronics housing: –40 °C T
a
+80 °C.
Observe the information in the temperature tables.
Basic specification, Position 5 (Housing) = C
In Zone 0, avoid sparks caused by impact and friction.
Basic specification, Position 5 (Housing) = A
Avoid electrostatic charging of the housing (e.g. friction, cleaning, maintenance, strong medium flow).
Device type FMP52, FMP55
A probe coated with non-conductive material can be used if avoiding electrostatic charging of the probe
(e.g. through friction, cleaning, maintenance, strong medium flow).
Safety instructions:
Installation
XA00634Fxx01
å 1
A
Zone 0, Zone 1
1 Tank; Zone 0, Zone 1
2 Electronic insert
3 Certified associated apparatus
4 Potential equalization line
5 Potential equalization
A
+
1
2
4
4
5
3