6.1.6 Special mounting conditions
Concrete silos
Installation, for example, into a thick concrete ceiling should be made flush with the lower
edge. Alternatively, the probe can also be installed into a pipe that must not protrude over
the lower edge of the silo ceiling. The pipe should be kept at a minimum length.
Installation suggestions see diagram.
100
4
100
4
mm(in)
ø80...150
(ø3.2...6)
1
3
2
A0014138
1 Metal sheet
2 Metal tube
3 Extension rod / Centering HMP40 (see "Accessories")
Note for installations with rod extension/center washer (accessories): Strong dust
generation can lead to build-up behind the center washer. This can cause an interference
signal. For other installation possibilities please contact Endress+Hauser.