Liquiline CM442/CM444/CM448 Electrical connection
Endress+Hauser 27
98i97i88i87i
98i97i88i87i
BN
WH
GN
YE
Sensor
BN
WH
GN
YE
Sensor
2DS
Ex-i
A0045659
 30 Sensors without additional supply voltage at sensor communication module type 2DS Ex-i
Intrinsically safe sensors for use in explosive atmospheres may only be connected to the
sensor communication module type 2DS Ex-i. Only the sensors covered by the certificates
may be connected (see XA).
5.3 Connecting additional inputs, outputs or relays
L
WARNING
Module not covered
No shock protection. Danger of electric shock!
Change or extend the hardware for the non-hazardous area: always fill the slots from left
to right. Do not leave any gaps.
If all of the slots are not occupied in devices for the non-hazardous area: always insert a
dummy cover or end cover in the slot to the right of the last module. This ensures that the
unit is shock-protected.
Always ensure shock protection is guaranteed particularly in the case of relay modules (2R,
4R, AOR).
Hardware for the hazardous area may not be modified.
If additional shields are required, connect them with PE centrally in the control cabinet via
terminal blocks supplied by the customer.