The following parameters affect the safety function and are automatically adapted by the device
when configuring higher-ranking parameters (known as application parameters). Values can not be
specified as the higher-ranking parameters may be freely configurable. This indirect setting is
permitted in increased safety mode. However, it is not permitted to change the parameters directly.
If these parameters were changed directly, only expert mode is available for selection in the
SIL/WHG confirmation.
Parameter Parameter name
Setup → Advanced setup → Level
Medium property
Max. draining speed liquid
Max. filling speed liquid
Blocking distance
Expert → Sensor → Distance
Dead time
Integration time
Max. integration time
Delta at integration time
Expert → Sensor → Safety settings
Echo lost window right
Echo lost window left
Draining speed
Filling speed
Expert → Sensor → Envelope curve
Envelope statistics down
Envelope statistics up
FAC offset
Expert → Sensor → Mapping
Map gap
End of mapping
Expert → Sensor → Echo fine adjustment Fine adjustment mode
Expert → Sensor → Tank bottom evaluation
Min. amplitude TBD
Lower level area
Expert → Sensor → Echo tracking
Window size tracking
Maximal track counter