Commissioning Gammapilot M FMG60 PROFIBUS PA
44 Endress+Hauser
*14
value full
[%]
target value
at calibr. point
*15
calibration
0: stop / edit
1: start
*11
avg.
puls rate
*16
full calibr.
[c/s]
-1: not
calibrated
*13
calibr. point
0: full/covered
1: empty/free
1
*17
value empty
[%]
target value
at calibr. point
*15
calibration
0: stop / edit
1: start
*11
avg.
puls rate
*18
empty calibr.
[c/s]
-1: not
calibrated
1
*19
next point
0: no
1: yes
Group
selection
E
0
E
0
A0037162-EN
The "Value full" (*14) and "Value empty" (*17) functions only appear if the "Level" option
was selected in the "Measurement mode" (*05) function.
"Calibration point" (*13)
Local display
Calibr. point *13
Full/covered
Empty/free
This function is used to select which calibration point ("full/covered" or "empty/free") will be