Proline Prowirl 72F, 72W, 73F, 73W
Endress+Hauser 11
Wiring (remote version) Note!
The remote version must be grounded. In doing so, the sensor and transmitter must be connected to the
same potential matching system.
A0001893
Connecting the remote version
a = Connection compartment cover (transmitter)
b = Connection compartment cover (sensor)
c = Connecting cable (signal cable)
d = Identical potential matching for sensor and transmitter
e = Connect shielding to ground terminal in transmitter housing and keep as short as possible
f = Connect shielding to ground terminal in connection housing
Wire colors (colour code according to DIN 47100):
Terminal number: 1 = white; 2 = brown; 3 = green; 4 = yellow, 5 = gray; 6 = pink; 7 = blue; 8 = red
Supply voltage HART:
Non-Ex: 12...36 V DC (18...36 V DC)
Ex i and Ex n: 12...30 V DC (18...30 V DC)
Ex d: 15...36 V DC (21...36 V DC)
PROFIBUS PA and FOUNDATION Fieldbus:
Non-Ex: 9...32 V DC
Ex i and Ex n: 9...24 V DC
Ex d: 9...32 V DC
Current consumption PROFIBUS PA: 16 mA, FOUNDATION Fieldbus: 16 mA
Cable entry Power supply and signal cables (outputs):
Cable entry M20 x 1.5 (8...11.5 mm)
Thread for cable entry: ½" NPT, G ½", G ½" Shimada
Fieldbus connector
Power supply failure Totalizer stops at the last value determined (can be configured)
All settings are kept in the EEPROM
Error messages (incl. value of operated hours counter) are stored