7.4.3 Sensor calibration
Sensor calibration adjusts the weight measurement of the detector unit. The calibration
consists of three steps as follows.
ADC zero calibration
ADC offset calibration
ADC span calibration
For the ADC offset weight calibration, either 0 g or an offset weight (0 to 100 g) can be used.
Using an offset weight other than 0 g is recommended for density measurement.