Operating Instructions
Proline Prosonic Flow B 200
HART
Ultrasonic transit time flowmeter
BA01031D/06/EN/03.15
71283621
Valid as of version
01.02.zz (Device firmware)
AUTHORIZED DISTRIBUTOR:
InstrumentsAndControl.com
Houston, Texas USA
sales@InstrumentsAndControl.com
832-615-3588

Other manuals for Endress+Hauser Proline Prosonic Flow B 200

Questions and Answers

  • S
    sabrina36Aug 17, 2025
    What to do if my Endress+Hauser Proline Prosonic Flow B 200 Measuring Instruments device measures incorrectly?
    • A
      Ashley HamiltonAug 17, 2025
      If your Endress+Hauser Measuring Instruments device is measuring incorrectly, it could be due to a configuration error or because the device is being operated outside of its intended application. First, check and correct the parameter configuration. Second, observe the limit values specified in the "Technical Data".
  • R
    Robin MilesAug 20, 2025
    Why is the signal output outside the valid range for my Endress+Hauser Measuring Instruments?
    • B
      Bryan WhiteAug 20, 2025
      If the signal output is outside the valid range for your Endress+Hauser Measuring Instruments, the main electronics module could be defective. In this case, you should order spare part ? ? 124.
  • F
    Frank MasonAug 22, 2025
    Why is the local display dark and there are no output signals on my Endress+Hauser Measuring Instruments?
    • J
      johnbeasleyAug 23, 2025
      The local display might be dark with no output signals on your Endress+Hauser Measuring Instruments due to several reasons: * The supply voltage doesn't match the nameplate specifications, so you should apply the correct supply voltage. * The supply voltage polarity is incorrect, thus you should reverse the polarity. * There's no contact between connecting cables and terminals, so check and correct the cable connections if needed. * The terminals aren't correctly plugged into the I/O electronics module, so check the terminals. * The I/O electronics module is defective, in which case you should order spare part ? ? 124.
  • J
    Jacob BennettAug 25, 2025
    What to do if the local display is dark but the signal output is within the valid range on my Endress+Hauser Measuring Instruments?
    • W
      wyoungAug 26, 2025
      If the local display is dark, but the signal output is within the valid range on your Endress+Hauser Measuring Instruments, it could be due to a few reasons: * The display is set too bright or too dark. To set the display brighter, simultaneously press ? + ?. To set the display darker, simultaneously press ? + ?. * The cable of the display module is not plugged in correctly. Insert the plug correctly into the main electronics module and display module. * The display module is defective. In this case, order spare part ? ? 124.
  • T
    Tom MartinAug 27, 2025
    How to fix no write access to parameters on Endress+Hauser Measuring Instruments?
    • W
      wileyalexanderAug 28, 2025
      If you have no write access to parameters on your Endress+Hauser Measuring Instruments, it might be due to: 1. Hardware write protection being enabled: Set the write protection switch on the main electronics module to the OFF position. 2. The current user role having limited access authorization: Check the user role ? ? 46 and enter the correct customer-specific access code ? ? 46.
  • L
    Leslie BriggsAug 30, 2025
    What does it mean if the backlighting of local display is red on Endress+Hauser Proline Prosonic Flow B 200 Measuring Instruments?
    • G
      Gerald WalkerAug 30, 2025
      If the backlighting of the local display is red on your Endress+Hauser Measuring Instruments, it indicates that a diagnostic event with "Alarm" diagnostic behavior has occurred. Take remedial measures ? ? 112.
  • L
    Lisa LopezSep 1, 2025
    How to resolve a "Communication Error" message on the local display of Endress+Hauser Proline Prosonic Flow B 200?
    • D
      duranmargaretSep 1, 2025
      If you see a "Communication Error" message on the local display of your Endress+Hauser Measuring Instruments, it means that the communication between the display module and the electronics is interrupted. Check the cable and the connector between the main electronics module and display module. If the problem persists, order spare part ? ? 124.
  • J
    Jessica CummingsSep 4, 2025
    How to resolve text on local display appearing in a foreign language and cannot be understood on Endress+Hauser Proline Prosonic Flow B 200?
    • G
      glensilvaSep 4, 2025
      If the text on the local display appears in a foreign language and cannot be understood on your Endress+Hauser Measuring Instruments, it means that an incorrect operating language is configured. To fix this: 1. Press ? + ? for 2 seconds ("home position"). 2. Press ?. 3. Set the desired language in the Language parameter.
  • A
    Alicia JohnsonSep 7, 2025
    How to fix no connection via HART protocol on Endress+Hauser Proline Prosonic Flow B 200?
    • B
      Brandy HernandezSep 7, 2025
      If there is no connection via HART protocol on your Endress+Hauser Measuring Instruments, it may be due to: * A missing or incorrectly installed communication resistor. Install the communication resistor (250 ?) correctly and observe the maximum load ? ? 26. * Issues with the Commubox, such as incorrect connection, configuration, driver installation, or USB interface configuration. Refer to the Commubox documentation ? FXA195 HART: Document "Technical Information" TI00404F.
  • T
    Timothy BradleySep 9, 2025
    Why is the signal output outside the valid current range (< 3.6 mA or > 22 mA) for my Endress+Hauser Proline Prosonic Flow B 200?
    • D
      Dorothy YangSep 9, 2025
      If the signal output is outside the valid current range (< 3.6 mA or > 22 mA) for your Endress+Hauser Measuring Instruments, the I/O electronics module is likely defective. Order spare part ? ? 124.