Technical data Proline Prosonic Flow B 200 HART
130 Endress+Hauser
16 Technical data
16.1 Application
The measuring device is suitable for flow measurement of gases only.
Depending on the version ordered, the measuring device can also measure potentially
explosive, flammable, poisonous and oxidizing media.
To ensure that the device remains in proper operating condition for its service life, use the
measuring device only for media against which the process-wetted materials are
adequately resistant.
16.2 Function and system design
Measuring principle Proline Prosonic Flow uses a measurement method based on the transit time difference.
Measuring system The device consists of a transmitter and a sensor.
The device is available as a compact version:
The transmitter and sensor form a mechanical unit.
For information on the structure of the device → 12
16.3 Input
Measured variable Direct measured variables
Volume flow
Calculated measured variables
• Corrected volume flow
• Mass flow
Optional measured variables (can be ordered)
Order code for "Sensor version", option 2 "Volume flow + Biogas analysis"
• Corrected methane volume flow
• Energy flow
• Methane fraction
• Gross calorific value
• Wobbe index
• Temperature
Measuring range Standard (order code for "Calibration flow", option 1 "Operable flow range 30 : 1")
Nominal diameter Velocity Effective volume flow
[mm] [in] [m/s] [ft/s] [m
3
/h] [ft
3
/h]
50 2 1 to 30 3.28 to 98.4 9 to 269 316 to 9 495
80 3 1 to 30 3.28 to 98.4 20 to 611 720 to 21 592
100 4 1 to 30 3.28 to 98.4 34 to 1 032 1 215 to 36 443