Diagnostics and troubleshooting Proline Prosonic Flow B 200 HART
114 Endress+Hauser
Diagnostic
number
Short text Remedy instructions Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
412 Processing download Download active, please wait C Warning
431 Trim 1 to 2 Carry out trim C Warning
437 Configuration
incompatible
1. Restart device
2. Contact service
F Alarm
438 Dataset 1. Check data set file
2. Check device configuration
3. Up- and download new
configuration
M Warning
441 Current output 1 to 2 1. Check process
2. Check current output settings
S Warning
1)
442 Frequency output 1. Check process
2. Check frequency output settings
S Warning
1)
443 Pulse output 1. Check process
2. Check pulse output settings
S Warning
1)
444 Current input 1 1. Check process
2. Check current input settings
S Warning
1)
452 Calculation error 1. Check device configuration
2. Check process conditions
S Alarm
1)
453 Flow override Deactivate flow override C Warning
484 Simulation Failure
Mode
Deactivate simulation C Alarm
485 Simulation measured
variable
Deactivate simulation C Warning
486 Simulation current
input 1
Deactivate simulation C Warning
491 Simulation current
output 1 to 2
Deactivate simulation C Warning
492 Simulation frequency
output
Deactivate simulation frequency
output
C Warning
493 Simulation pulse
output
Deactivate simulation pulse output C Warning
494 Switch output
simulation
Deactivate simulation switch output C Warning
495 Diagnostic event
simulation
Deactivate simulation C Warning
Diagnostic of process
801 Supply voltage too low Increase supply voltage S Warning
1)
803 Current loop 1. Check wiring
2. Change I/O module
F Alarm
832 Electronic temperature
too high
Reduce ambient temperature S Warning
1)
833 Electronic temperature
too low
Increase ambient temperature S Warning
1)
834 Process temperature
too high
Reduce process temperature S Warning
1)
835 Process temperature
too low
Increase process temperature S Warning
1)
836 Process pressure Reduce process pressure S Alarm
837 Process pressure Increase process pressure S Warning
1)