Proline Prosonic Flow B 200 HART Diagnostics and troubleshooting
Endress+Hauser 115
Diagnostic
number
Short text Remedy instructions Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
841 Sensor range Check flow velocity S Warning
1)
842 Process limit Low flow cut off active!
1. Check low flow cut off
configuration
S Warning
881 Sensor signal path 1 to
2
1. Check process conditions
2. Clean transducer
3. Change transducer
M Warning
1)
882 Input signal 1. Check input configuration
2. Check external device or process
conditions
F Alarm
930 Process fluid Sound velocity too high! Check process
conditions
S Warning
931 Process fluid Sound velocity too low! Check process
conditions
S Warning
1) Diagnostic behavior can be changed.
12.6 Pending diagnostic events
The Diagnostics menu allows the user to view the current diagnostic event and the
previous diagnostic event separately.
To call up the measures to rectify a diagnostic event:
Via local display →  109
Via "FieldCare" operating tool →  111
Other pending diagnostic events can be displayed in the Diagnostic list submenu
→  116
Navigation
"Diagnostics" menu
Structure of the submenu
Diagnostics
Actual diagnostics
Previous diagnostics
Operating time from restart
Operating time