Technical alterations reserved
Application
conditions
(continued)
EMC immunity
RF protection To EN 55011 Group 1, Class A
Interferance safety
ESD To EN 61000-4-2, 6 kV/8 kV
Electromagnetic
fields
To EN 61000-4-3, 10 V/m
Burst (supply) To EN 61000-4-4, 2 kV
Surge To EN 61000-4-5, 1 kV
Cable high
frequency
To EN 61000-4-6, 10 V
Mechanical
construction
Dimensions H: 48 mm, W: 96 mm, D: 90 mm
Weight 300 g
Materials Housing front: Die cast aluminium
Housing casing: Galvanised sheet steel
Housing rear panel: Plastic ABS
Electrical
connection
Plug on screw terminals (fixable), Terminal size
1.5 mm² solid core, 1.0 mm² stringed with ferrules
Display and
operating level
Display 5 digit LC display, 17 mm character size
Display range - 19999 to + 99999
Offset - 19999 to + 32767
Operation 3 push button operation (-/+/E)
Power supply
Power supply Via 4...20 mA current loop
Volt drop < 2 Volt
Certification
CE Directives 89/336/EWG
Ex certification ATEX II 1 G EEx ia IIC T6
FM IS, Class I, Div. 1+2, Group A,B,C,D
CSA IS, Class I, Div. 1+2, Group A,B,C,D
CSA GP CSA General purpose
Process display
28