EasyManua.ls Logo

Endress+Hauser RIA251 - Page 30

Endress+Hauser RIA251
88 pages
Go to English
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Technical alterations reserved
Application
conditions
(continued)
EMC immunity
RF protection To EN 55011 Group 1, Class A
Interferance safety
ESD To EN 61000-4-2, 6 kV/8 kV
Electromagnetic
fields
To EN 61000-4-3, 10 V/m
Burst (supply) To EN 61000-4-4, 2 kV
Surge To EN 61000-4-5, 1 kV
Cable high
frequency
To EN 61000-4-6, 10 V
Mechanical
construction
Dimensions H: 48 mm, W: 96 mm, D: 90 mm
Weight 300 g
Materials Housing front: Die cast aluminium
Housing casing: Galvanised sheet steel
Housing rear panel: Plastic ABS
Electrical
connection
Plug on screw terminals (fixable), Terminal size
1.5 mm² solid core, 1.0 mm² stringed with ferrules
Display and
operating level
Display 5 digit LC display, 17 mm character size
Display range - 19999 to + 99999
Offset - 19999 to + 32767
Operation 3 push button operation (-/+/E)
Power supply
Power supply Via 4...20 mA current loop
Volt drop < 2 Volt
Certification
CE Directives 89/336/EWG
Ex certification ATEX II 1 G EEx ia IIC T6
FM IS, Class I, Div. 1+2, Group A,B,C,D
CSA IS, Class I, Div. 1+2, Group A,B,C,D
CSA GP CSA General purpose
Process display
28

Table of Contents