EasyManua.ls Logo

Endress+Hauser tmt142 - Terminal assignment

Endress+Hauser tmt142
48 pages
Go to English
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Wiring TMT142 HART®
24
8. Make sure that the terminal screws are tight. Re-seal the cable gland by screwing the cover
back on. In doing so, also pay particular attention to Chap. 3.4.
9. On completion of the wiring, screw on the electronics module (item 4) again, plug on the
display with twist protection (item 3) again if applicable and screw the housing cover
together with the O-ring (item 2) down again. Finally refit the cover clamp (item 1).
In order to avoid connection errors always take note of the hints given in the section
connection check!
3.1 Quick wiring guide
3.1.1 Terminal assignment
A0026193-EN
Fig. 3: Wiring the field transmitter - terminal assignment
It is possible to adjust the display with retainer in 90°stages to your desired position
and rearrange it on the particular slot in the electronics module. It is not possible to
rotate the display when the display is plugged on already.
3
3
3
3
3
3
1
1
1
1
1
1
1
4
4
2
2
2
2
2
RTD
W
RTD
W
RTD
W
TC
+
-
1
2
3
4
+
-
3
4
S1
-
+
#
4-wire
2-wire 3-wire
11 to 40 V
11 to 30 V Ex
4 to 20 mA
HART® - Signal
Sensor
Sensor
ESD - electrostatic discharge.
Protect the terminals from electrostatic discharge. Failure to observe this may result in
destruction or malfunction of parts of the electronics.