49
Surge Program : on MF and DF
Polarity : Positive or Negative
Amplitude : 0 - 70 mA
MF interrupted galvanic current
Frequency : 8000 Hz
Duty cycle : 90%
Polarity : Positive or Negative
Amplitude : 0 - 40 mA
Direct Galvanic current
Polarity : Positive or Negative
Amplitude : 0 – 40 mA
Faradic Rectangular Pulsed Current
Phase Duration : 0.02 – 1000 ms
Pulse Frequency : 0.2 – 1 Hz in steps of 0.1 Hz, 1 – 200 in steps of 1 Hz
Surge Program : Yes
Polarity : Positive or Negative
Amplitude : 0 - 80 mA
Faradic Triangular Pulsed Current
Phase duration : 0.1 – 1000 ms
Pulse Frequency : 0.2 – 1 Hz in steps of 0.1 Hz, 1 – 200 in steps of 1 Hz
Surge program : Yes
Polarity : Positive or Negative
Amplitude : 0 - 80 mA
Träbert 2-5 (Rectangular Pulsed Current)
Phase duration : 2 ms
Interval : 5 ms
Polarity : Positive or Negative
Amplitude : 0 - 80 mA
10.3 Technical Data
Mains voltage : 100 – 240 V ± 10%
Mains frequency : 50 / 60 Hz
Main unit:
Dimensions stand alone : 24 x 32 x 12 cm (w x d x h)
Dimensions on inclination foot : 24 x 30.5 x 18.2 cm (w x d x h)
Dimensions on Vacotron : 24 x 30.5 x 21.6 cm (w x d x h)
Weight : 2 Kg
Weight including optional battery : 3 Kg
Vacotron:
Dimensions : 24 x 28.6 x 9.3 cm (w x d x h)
Weight : 2 Kg
Environmental conditions for transport and storage:
Temperature : -20° to +70° C
Relative humidity : 10 to 90 % (contained in original packaging)
Atmospheric pressure : 500 to 1060 hPa
Environmental conditions for normal use:
Temperature : 10° to 40° C
Relative humidity : 10 to 90 % non condensing
Atmospheric pressure : 500 to 1060 hPa
Technical modifications reserved