Sample Conditioning System Cleaning Procedure
During start-up or plant upset situations, the 330S/331S H
2
S analyzer may become
contaminated with amine or hydrogen sulfide scavenger solution. This may cause the
analyzer to read low (this can be determined at calibration). If the analyzer reads low,
it will require incremental increases in the gain to maintain calibration. Please refer to
factory calibration sheet for factory set gain factor. The scavenger solution is water
soluble and therefore is relatively easy to clean.
Material List
Cleaning Kit Part Number: 330900
Alconox RBS Solid, powdered precision cleaner w/ MSDS (2.5 tbsp)
330079 – Rear window and gasket
MIF-225-PP – Small nipple (x2)
MIF-32-PVC – Grey nipple
MIF-32-PP – Big nipple
MIF-24-PVC – Humidifier replacement tube fittings (x2)
MIP-4 Humidifier replacement plug
06349-40 – Black 1/8" male/female elbow
LT-2-4 (1/4" x 11") – Tubing for sample chamber to humidifier
LT-2-4 (1/4" x 10") – Tubing for flow meter to humidifier
LT-3-5 (3/16" x 10") – Tubing for sample chamber to vent block
PFA (24") – Tubing for flowmeter to bulkhead
Do not use solvents, brake cleaners, soaps, detergents or rubbing alcohol to clean up
analyzer or sample system.
Procedure
1. Mix a 1% (2-1/2 tbsp per gallon) of Alconox in warm water
2. Sample line tubing
Shut off flow at the sample point prior to sample conditioning system
Flush the sample line and components with cleaning solution
Rinse with fresh water
Flush with isopropyl alcohol
Dry with clean, dry instrument air or gas
3. Sample conditioning system – Take pictures of SCS before disassembling
Remove filter elements from filter housings and discard