EasyManua.ls Logo

Envent 331S - Sample Conditioning System Issues

Envent 331S
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Envent Engineering Ltd. Page 42 Revision 2.4
Electronics Issues (Cont’d)
Possible Reasons
Possible Solutions
The system variable for output
has been modified.
By factory configuration, the analog outputs
(1 & 2) are configured to output based on
H2S Sample 1. Make sure that if they are
modified, that the right system variables are
selected.
Not coming out
of alarm
Alarms are latched If alarms are latched, they need to be
acknowledged. Go to the display> cycle
through until "ACK" is reached> press the
right button to acknowledge all latched
alarms.
To deactivate the latching on any alarm,
use the ICE software, and connect to the
analyzer and de-select latching on any
alarm that is latched.
The analyzer is actually in
alarm
Make sure the alarm setpoint values are as
desired and that the analyzer is below (or
above – like temperature setpoint) those
setpoint values.
Sample Conditioning System Issues
Possible Reasons
Possible Solutions
Liquid Carried
over in SCS
Sample too wet for current
conditioning sample system.
If a one-time occurrence: Cleaning required
for sample system, refer to "Sample
Conditioning System Cleaning Procedure"
on page 32.
If more than one time occurrence: Sample
conditioning system may need a system for
wet/dirty sample gas (extra filters set as
coalescing, add liquid float stops, etc.).
Consult Envent Engineering Ltd.
Regulator not
maintaining 15
psi (erratic H2S
readings)
Problems with the Regulator
(over pressured)
Replace Regulator, Consult Envent
Engineering Ltd.
Liquid carried over into
regulator
Consult Envent Engineering Ltd.
Problems with the (50 psi)
sample pre-regulator at the
sample point.
Heated regulator may be required if liquid
hydrocarbon carry over is present at the
pre-regulation sample.
Continued on next page

Table of Contents