Centrifuge 5804/5804 R and 5810/5810 R
3 Error lists / Troubleshooting
3-6 Error lists/Troubleshooting
Error 13 EDP error
Serial PC interface: reception error, parity
error, wrong code number, check sum
error.
Check the interface PCB and the plug
connections for correct positioning.
Error 14 EPROM check sum error after mains on.
– Error program memory, EPROM
defective.
No further actions possible.
– Replace the EPROM or the
PROCESSOR (IC1).
– Replace the "Control" PCB.
Error 15
Rotor
Imbalance error at n > 2000 1/min
Device brakes with normal ramp; device
was pushed or rotor not fixed.
–refer to ERROR 2.
– Rotor not fixed tightly.
– Device was pushed.
– Table is not stable.
Error 16 Error during request to FC.
– FC not ready, drive relay has dropped.
– Error after telegram repetition.
Device switches off – stops without
braking.
– Customer service program 9
• view the error table of the
converter.
– Check cable and connections at plug
X9 and X2.
– Check the overtemperature sensor of
the motor and its connections to the
converter.
Display Cause Remedy