Centrifuge 5804/5804 R and 5810/5810 R
3 Error lists / Troubleshooting
3-6 Error lists/Troubleshooting
Error 13 EDP error
Serial PC interface: reception error, parity
error, wrong code number, check sum
error.
Check the interface PCB and the plug
connections for correct positioning.
Error 14 EPROM check sum error after mains on.
Error program memory, EPROM
defective.
No further actions possible.
Replace the EPROM or the
PROCESSOR (IC1).
Replace the "Control" PCB.
Error 15
Rotor
Imbalance error at n > 2000 1/min
Device brakes with normal ramp; device
was pushed or rotor not fixed.
–refer to ERROR 2.
Rotor not fixed tightly.
Device was pushed.
Table is not stable.
Error 16 Error during request to FC.
FC not ready, drive relay has dropped.
Error after telegram repetition.
Device switches off – stops without
braking.
Customer service program 9
view the error table of the
converter.
Check cable and connections at plug
X9 and X2.
Check the overtemperature sensor of
the motor and its connections to the
converter.
Display Cause Remedy
MSTM/0408