Error lists/Troubleshooting 3-9
Centrifuge 5804/5804 R and 5810/5810 R
3 Error lists / Troubleshooting
Error 24
Clear
Memory
Program memory is full. Erase program if required,
or start new run with Start or Prog.
Error 25
Interrupt
Interrupt.
Mains interrupted during a run.
Is mains voltage generally too low or too
high?
Do deviations exist?
Cable in the device is damaged, e.g. after
replacement of the motor cap/rotational
speed sensor or PCB imbalance
detection.
–Restart.
– Check mains supply (during start of
cooling aggregate as well).
– Check cables in the device and look
for cable damage, particularly near the
motor.
Error 31 – x 0: Timeout while closing the lid.
1: Timeout while opening the lid.
2: Undefined position when opening.
3: Monitoring of microswitches for lid.
4: Released lid does not open.
5: Error during lid test after power-on.
– Check the connections on the sensor
board and plug X10.
Display Cause Remedy
MSTM/0408