Error lists/Troubleshooting 3-9
Centrifuge 5804/5804 R and 5810/5810 R
3 Error lists / Troubleshooting
Error 24
Clear
Memory
Program memory is full. Erase program if required,
or start new run with Start or Prog.
Error 25
Interrupt
Interrupt.
Mains interrupted during a run.
Is mains voltage generally too low or too
high?
Do deviations exist?
Cable in the device is damaged, e.g. after
replacement of the motor cap/rotational
speed sensor or PCB imbalance
detection.
–Restart.
Check mains supply (during start of
cooling aggregate as well).
Check cables in the device and look
for cable damage, particularly near the
motor.
Error 31 – x 0: Timeout while closing the lid.
1: Timeout while opening the lid.
2: Undefined position when opening.
3: Monitoring of microswitches for lid.
4: Released lid does not open.
5: Error during lid test after power-on.
Check the connections on the sensor
board and plug X10.
Display Cause Remedy
MSTM/0408