Centrifuge 5804/5804 R and 5810/5810 R
3 Error lists / Troubleshooting
3-10 Error lists/Troubleshooting
Errors from the service programs
Display
Cause
Error 32 Hardware error: imbalance measuring
system.
Check the connections on the sensor
board and plug X10.
Error 33 Measuring values outside of the
measuring range, too little space between
sensor and magnet.
Check the connections on the sensor
board and plug X10.
Error 34 Faulty measuring series.
Mean value not within limits or
distance too short.
Scattering run 1–3 too great.
Scattering run 4–6 too great.
Distance sensor – magnet too great.
Rotor is unbalanced during imbalance
adjustment.
Motor continues to vibrate.
Check the connections on the sensor
board and STV X10.
–See Error 2.
Adjust distance Sensor – Magnet,
see Sec. 4.
Clean pegs and nuts of the rotor/
bucket and grease slightly with
Eppendorf pivot grease.
Test another rotor.
When using Rotor A-4-81:
use magnet carrier imbalance
detection with 9x13 mm magnet.
Speed sensor cable jammed.
Replace shock mounts for motor.
Perform imbalance calibration on a
solid bench.
Error 35 Imbalance sensor idle value exceeds the
limit (670 ± 100).
Check the connections on the sensor
board and plug X10.
–see Error 34
MSTM/0408